Sulfur Compound Instruments
​​​​​​​​​​​​​​​​​​​ Conta​​​ct us
Traini​ng
Sales Contacts
Request a Quote
Terms & Conditions
​​ ​
​​

 

​Model T100H

with NumaView™ Software

High Range UV Fluorescence SO2 Analyzer

 

 Features

 

  • Ranges: 0-10 ppm to 0-5,000 ppm, user selectable
  • Dual ranges
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Fast response (<30 seconds to 95%)
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty

 Maintenance

 

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

 t100Hmaintenance.PNG

1 These items are required to maintain full warranty; all other items are strongly recommended.

2 A pump rebuild kit is available from Teledyne API's Technical Support, including all instructions and required parts.

 3 Replace desiccant bags each time the inspection plate for the sensor assembly is removed.

* There is no Internal IZS offered for the T100H.

 

 Downloads

  
  
  
06807G - T100 Manual.pdfT100 Manual7608 KB
T SERIES_6PG.BROCHURE_2018_web.pdfT Series with NumaView Brochure9567 KB
T100H Addendum_07265.pdfT100H Manual Addendum2649 KB
T100H Sales Literature.pdfT100H Specification944 KB

 Specifications

​​
Ranges Min: 0 - 10 ppm full scale
Max: 0 - 5,000 ppm full scale (selectable with dual range supported)
Measurement Units​ ppm, mg/m3 (selectable)
Zero Noise 0.1 ppm (RMS)
Span Noise < 1% of reading (RMS) above 10 ppm
Lower Detectable Limit 0.2 ppm
Zero Drift < 1 ppm/24 hours
Span Drift < 0.5% of full scale/24 hours
Lag Time 5 seconds
Rise/Fall Time < 30 seconds to 95%
Linearity 1% of full scale
Precision 0.5% of reading above 10 ppm
Sample Flow Rate 700 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight



Specifications are subject to change without notice.
All specifications are based on constant conditions.​

Analyzer: 31 lbs (16 kg)
External pump: 15 lbs (7 kg) ​

 Description

The Model T100H SO2 analyzer uses the proven UV fluorescence principle with the high measurement ranges and interferent rejection​​​ needed for extractive source monitoring.  The Model T100H provides a simple but precise solution for a broad range of CEM and stack testing applications.

The instrument design provides minimal quenching by O2 or CO2 and less than a 0.1% response to water in the sample. Additionally, special optical filtering reduces the effects of NO interference.​

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T100H comes with NumaView ™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView ™ Software. ​