​Model T703
with NumaView™ Software

​Photometric Ozone Calibrator

 

 Features

  • Primary or transfer standard for calibration of ozone analyzers
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 12 independent timers for sequences
  • Nested sequences (up to 5 levels)
  • UV Lamp Feedback modes: current control; photometer control
  • Internal zero air pump (optional)
  • Flow adjustable from 1 LPM to 5 LPM
  • High Flow/High Output capability (optional)
  • Inlets for external ozone reference sources
  • Two-year warranty

 Downloads

  
  
  
083730703B - MANUAL, USER, NVS, T703-T703U.pdfT703/T703U with NumaView™ Software Manual12066 KB
ESD Fundamentals - 04786.pdfFundamentals of Electro-Static Discharge (ESD)216 KB
NVS Remote Manual - 08492.pdfNumaView™ Software Remote User Guide867 KB
SAL000064D - T703.pdfT703 Specification1147 KB
SAL000064E-CN T703.pdfT703规格 (Chinese Specification)5230 KB
SAL000076K - Environmental Instruments Overview.pdfEnvironmental Instruments Overview610 KB
T SERIES_6PG.BROCHURE_2018_web.pdfT Series with NumaView™ Software Brochure9567 KB

 Specifications

​​
​​System
Linearity
± 1% of full scale
Precision​ 1.0 ppb
Response Time < 180 seconds to 95%
Stability (7 days) < 1% with photometer feedback
< 3% without photometer feedback (CNST or REF)
​​Ozone Generator Module
Flow Rate (with internal zero air source) 1 to 5 LPM adjustable
Flow Rate (with external zero air source) 1 to 15 LPM adjustable
Maximum Output 5 LPM
Minimum Output 100 ppb LPM
Maximum Concentration 5 ppm at 1 LPM
Minimum Concentration 50 ppb at 2 LPM
Response Time < 180 seconds to 98%​
​​UV Photometer
Range 0 - 100 ppb to 0 - 10 ppm (selectable)
Precision 1.0 ppb
Linearity ± 1% of full scale
Rise/Fall Time < 20 seconds to 95% (photometer response)
Response Time < 180 seconds to 95% (system response)
Zero Drift < 1.0 ppb/7 days
Span Drift < 1%/24 hours
Lag Time < 10 seconds
Flow Rate 800 cc/min ±10%
​​Electrical Specifications
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges (Test Channel) 10V, 5V, 1V, 0.1V (selectable)
​​Communications Specifications
Included I/O

1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
12 x digital control outputs
12 x digital control inputs
8 x digital status outputs
Optional I/O 1 x USB com port
1 x RS485
Multidrop RS232
Physical Specifications
Operating Temperature Range 5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 24” (178 x 432 x 609 mm)
Weight


Specifications are subject to change without notice.
All specifications are based on constant conditions.​

35.5 lbs (16.1 kg) with internal zero air pump ​​

 Description

 

The Model T703 is a microprocessor controlled ozone calibrator for precision gas analyzers.  ​The Model T703 can deliver concentrations from 50 ppb to 5 ppm at flows from 1 LPM to 5 LPM.

The Model T703 has two major components; the ozone generating device and the photometer.  The ozone generator produces ozone by passing zero air over a UV lamp.  Flow rate and lamp intensity are adjusted to deliver a specified ozone concentration by volume.  The temperature and pressure compensated photometer, identical to that of the Model T400 ozone analyzer, provides feedback control of the UV lamp intensity, assuring stable ozone output.  A nearly unlimited number of calibration sequences may be programmed into the non-volatile memory of the T703, covering time periods up to a year.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T703 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.