Oxygen Compound Instruments
​​​​​​​​​​​​​​​​​​​ Conta​​​ct us
Traini​ng
Sales Contacts
Request a Quote
Terms & Conditions
​​ ​
​​

 

​Model T265
with NumaV​iew™ Software

Chemiluminescence 03 Analyzer

​​


 Features

 

  • Ranges: 0-100 ppb to 0-2,000 ppb, user selectable
  • Independent ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty
  • US EPA Federal Reference Method (RFOA-0216-230)

      View the US EPA Proposed FRM Report

 Maintenance

​The following table provides a list of items that need regular maintenance to ensure your product remains in good working condition.

ITEMACTIONFREQUENCYCAL CHECK REQUIRED
TEST functionsReview and evaluateWeeklyNo
Particulate filterChange particle filterWeekly (if in stack system: As Needed)No
Zero/span checkEvaluate offset and slopeWeeklyNo
Zero/span calibrationZero and span calibrationEvery 3 monthsYes
External zero air scrubber optionExchange chemicalEvery 3 monthsNo
External dryer optionReplace chemicalWhen indicator color changesNo
Ozone cleanserChange chemicalAnnuallyYes
Reaction cell window (“optical filter”  Figure 11‑6 in manual)CleanAnnually or as necessaryYes
DFU filtersChange particle filterAnnuallyNo
Pneumatic sub-systemCheck for leaks in gas flow pathsAnnually or after repairs involving pneumaticsYes if a leak is repaired
Reaction cell O-rings & sintered filtersReplaceAnnuallyYes 
PMT Sensor Hardware CalibrationLow-level hardware calibrationOn PMT/ preamp changes or if slope is outside of 1.0±0.3Yes
PumpRebuild headwhen RCEL pressure exceeds 10 in-Hg-A (at sea level)Yes
Inline Exhaust ScrubberReplaceAnnuallyNo
NO2 converterReplace converterEvery 3 years or if conversion efficiency drops below 96%Yes 
Desiccant  bagsReplaceAny time PMT housing is opened for maintenancen/a
 ​  ​

 Downloads

  
  
  
SAL000052E - T265.pdfT265 Specification1218 KB
T SERIES_6PG.BROCHURE_2018_web.pdfT Series with NumaView Brochure9567 KB
T265 Manual Addendum_07337.pdfT265 Manual3684 KB

 Specifications

Ranges Min: 0-100 ppb full scale
Max: 0-2,000 ppb ​full scale (selectable, dual range supported)
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.15 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 100 ppb
Lower Detectable Limit < 0.3 ppb
Zero Drift < 0.5 ppb/24 hours
Span Drift < 0.5% of full scale/24 hours
Lag Time < 10 seconds
Rise/Fall Time < 20 seconds to 95%
Linearity < 1% of full scale
Precision < 0.5% of reading
Sample Flow Rate 500 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4-20mA current outputs
Operating Temperature Range 5 - 40ºC (with US EPA Approval)
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight 40 lbs (18.1 kg)
Certifications


Specifications are subject to change without notice.
All specifications are based on constant conditions.​
US EPA Federal Reference Method: RFOA-0216-230
US EPA Federal Equivalent Method: EQOA-0611-199​​​​

 Description

The Model T265 is the only commercially available US EPA Federal Reference Method for Ambient Ozone Monitoring according to the latest US EPA regulatory rules for O3. The instrument uses Chemiluminescence with Nitric Oxide (NO) as a reagent, making it safe, stable, and effective for all applications. The NO-CL method for Ozone has several advantages over the conventional UV absorption technology, including the avoidance of positive and commonly found interferrents such as VOCs, fine particulates, Mercury, turpines, ClO2 and others. If these pollutants are present in the atmosphere, it can cause UV absorption analyzers to give an artificially high ozone measurement. The innovative T265 combines speed of response, sensitivity and stability of our proven nitrogen oxide analyzer, with a simple user interface, to provide reliable ozone measurements under the harshest conditions.

An inexpensive external cylinder of NO provides over one year of reagent for the T265.  A unique AutoZero feature provides superb stability by continuously correcting for zero drift, while advanced Adaptive Filtering allows the analyzer to optimize performance under changing conditions.  Optional zero and span valves allow automatic, unattended calibration checks.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T265 comes with NumaView™ Software.  NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.