The Model 701 is an excellent source of clean, dry air for dilution calibrators. It also may be used as a source of purge air for permeation tube ovens or burner air for FID analyzers. The regenerative, heatless dryer removes water and produces a gas with a dewpoint of less than -20°C, independent of inlet dewpoint. The regenerative dryer also assists in the removal of other gases, greatly increasing the life of the chemical scrubbers. The basic Model 701 includes an oil-less diaphragm-less pump plus scrubbers to remove SO2, NO, NO2, O3 and H2S. Optional scrubbers are available to remove CO and Hydrocarbons. Inlet air is pulled into the pump and routed through a pre-cooler and water trap, which removes water. The air then passes through the regenerative scrubber, where the final drying occurs, and from there to the storage tank. Tank pressure is monitored and maintained at a preset level by cycling the pump automatically as needed, extending pump and scrubber life. The outlet air from the tank is cleaned and goes through a filter to produce a clean and dry zero air supply. A microcontroller cycles the regenerative dryer and the water trap valves, ensuring that the pump does not start against full head pressure. The Model 701 is an ideal accessory to the Model 700 Mass Flow Calibrator or Model 702 Single Point Calibrator as a zero air source for analyzers. Features
|
|||||||||||||||||||||||||||||||
|