Home    Company Profile    What's New    Contact Us   

 
Model 454

Download the brochure

Process Ozone Monitor

The Model 454 is a microprocessor based gas monitor for measuring mid to high concentrations of ozone in oxygen or air. Available in a standard DIN panel mount
configuration with a 4-digit LED display, the compact design provides ease of mounting and installation.

The 454 Series is available in full scale ranges from 0-5% w/w to 0-25% w/w, and is factory calibrated in accordance with International Ozone Association standards. The measurement principle is based on the
absorption of UV light at 254nm. As the sample containing ozone flows through the absorption cell, it is
exposed to UV light at the 254nm wavelength. The amount of absorption by ozone is measured by a narrow bandwidth UV detector.

A reference detector measures the lamp output to compensate for changes in light intensity. This method, using a narrow spectrum UV source and a narrow bandwidth detector, provides excellent performance and stability without requiring additional UV filters.

The 454 Series’ temperature and pressure compensation normalizes readings to weight/volume. A front panel keypad displays concentration or cell pressure, while LEDs indicate monitoring status. The units employ a state-of-the-art microprocessor design to provide accurate, dependable monitoring of the ozone process.

Features
  • Standard one year warranty
  • Ranges 0-5% w/w to 0-25% w/w
  • DIN panel mount
  • High visibility, 4-digit display
  • Temperature and pressure compensation
  • Units in % w/w or g/Nm3
  • Narrow bandwidth detector eliminates need for UV filters
  • Analog and RS232/485 outputs
  • Status LEDs to monitor performance
  • 316 stainless steel cell
Specifications
Ranges: 0 - 5, 0 - 10, 0 - 15, 0 - 20% w/w,
0 - 25% w/w; 0 - 25, 0 - 100, 200, 300,
400 g/Nm3
Pressure and temperature compensation: Standard
Inlet pressure: 7 to 30 psia
Proof pressure: 100 psia
Repeatability: 0.5% of full scale
Response time: <30 seconds to 95%
Zero drift: 1% Full Scale/month (non-cumulative)
Operating temperature range: 5 - 45° C
Wetted materials: 316 Stainless Steel, PTFE, Sapphire, AlO3
Fittings: 1/4" stainless steel
Flow: 0.5 to 2.0 LPM
Power: 90-265 VAC, 50/60 Hz
Display: 4 digit LED readout
Analog output: 0–5V Full Scale; 4-20mA or 0-20mA isolated outputs optional
Zero calibration: Front panel keypad, contact closure input
Digital Outputs: Sensor OK, Lamp Low, Cell Dirty, Invalid Reading,    (Opto-isolated)
Serial data Interface: RS485 or RS232
Dimensions (HxWxD): 3.83” (97 mm) x 7.61” (193 mm) x 7.63” (194 mm)
Weight: 4.0 lbs. (1.8 kg)


   

Home | Terms of Use
Terms and Conditions - (PDF)
e-mail us: api-sales@teledyne.com

Copyright © 2008 Teledyne Technologies Incorporated. All rights reserved.