Problem Report Forms

Model 201E - Warranty Repair Questionnaire

A PDF version of this form is available here.

Company: *
Contact name: *
Telephone: *
Fax:
Email: *
Serial #:
Software Version:
Describe the Failure Symptoms:
List All Warning Messages:
Test values:
ParameterObserved valueUnitsNominal Range
Range PPB/PPM 50 PPB to 2000 PPB
NH3 Stab PPB/PPM < =1 PPB with Zero Air
Sample Flow CM3 500 ± 50
Ozone Flow CM3 80 ± 15
PMT Signal w/ Zero Air mV -20 to 150
PMT Signal at Span Gas Conc mV
PPB
0-5000 mV
0-20,000 PPB
Norm PMT Signal at Span Gas Conc mV
PPB
0-5000 mV
0-20,000 PPB
AZero mV -20 to 150
HVPS V 400 - 900
RCELL Temp °C 50 ± 1°C
Box Temp °C Ambient ± 5°C
PMT Temp °C 7 ± 2°C
Block Temp °C 50 ± 1°C
MOLY Temp °C 315 ± 5°C
RCEL Press in-Hg-A < 10
Sample Pressure in-Hg-A 3 to 5 <in. Ambient Absolute
TNx Slope  1.0  ± 0.3
TNx Offset  -50 to +150
NOx Slope  1.0  ± 0.3
NOx Offset  -50 to +150
NO Slope  1.0  ± 0.3
NO Offset  -50 to +150
ETEST PMT mV 2000 ±1000
OTEST PMT mV 2000 ± 1000
NH3 Converter Temp °C 825 ± 5°C
Values are in the Signal I/O
REF_4096_MV mV 4096mV ± 2 mV and Must be Stable
REF_GND mV 0 ± 0.5 and Must be Stable
What is the RCELL and sample pressure with the sample inlet on rear of machine capped?
RCELL: in-Hg-A Sample Pressure: in-Hg-A
If possible, please include a portion of the strip chart pertaining to the problem. Circle pertinent data.