Problem Report Forms

Model 200AU - Warranty Repair Questionnaire

A PDF version of this form is available here.

Company: *
Contact name: *
Telephone: *
Fax:
Email: *
Serial #:
Software Version:
Describe the Failure Symptoms:
List All Warning Messages:
Test values:
ParameterObserved valueUnitsNominal Range
Rangeppb5 - 2,000
NOx Stabilityppb<1
Sample Flowcm³/min1000 ± 100 (Std)
Ozone Flowcm³/min80 ± 15
Norm PMT@ZeromV0 - 100
Norm PMT@SpanmV0 - 5,000
Span Conc.ppb 
PREREACTmV0 - 1,000
HVPSV450 - 900 constant
DCPSmV2,500 ± 200
Rcell TempºC40 ± 1
Box Temp.ºCAmb ± 5
PMT Temp.ºC-5 ± 1
BLOCK Temp.ºC 40 ± 1
Moly Temp.ºC315 ± 5
Rcell Press.In-Hg-Aconstant 1-4
Samp. Press.In-Hg-Aconstant 25-30
NO Slope 1.0 ± 0.3
NOx Slope 1.0 ± 0.3
NO Offset -10 - +50
NOx Offset -10 - +50
E TestmV2,000 ± 500
O TestmV2,000 ± 1,000
Is unit leak checked?Yes No
What is the value of the gas in your span bottle?
Are you diluting your bottle?
What are you using for zero air?
Have you measured the sample flow? (if not do so)Yes No
Enter code to confirm submission
Validation Enter Confirmation Code:    *
 